Analysis of Percentage Depth Dose for 6 and 15 MV Photon Energies of Medical Linear Accelerator With CC13 Ionization Chamber

The success of external beam radiotherapy is mainly depended on how to commission and safety precaution is maintained before beam is heated to infected cells. CC13, an intermediate size ionization chamber is frequently used for the commissioning process of dose calculation for both standard and no...

Whakaahuatanga katoa

Ngā taipitopito rārangi puna kōrero
Kaituhi matua: Meaze, A. K. M. M. H.
Hōputu: Tuhinga
Reo:Ingarihi
I whakaputaina: NUCLEAR SCIENCE AND APPLICATIONS 2025
Ngā marau:
Urunga tuihono:https://repository.auw.edu.bd/handle/123456789/920
_version_ 1857448842604052480
author Meaze, A. K. M. M. H.
author_facet Meaze, A. K. M. M. H.
author_sort Meaze, A. K. M. M. H.
collection institutional Repository
description The success of external beam radiotherapy is mainly depended on how to commission and safety precaution is maintained before beam is heated to infected cells. CC13, an intermediate size ionization chamber is frequently used for the commissioning process of dose calculation for both standard and non-standard sizes field. The measurements of PDDs (Percentage of Depth Dose) were performed using TG-51 at the North East Cancer Hospital, Bangladesh using Varian Clinac IX-5982 with 6 MV & 15 MV photon beam energies for a set of 9 field sizes (4 × 4, 6 × 6, 8 × 8, 10 × 10, 15 × 15, 20 × 20, 25 × 25, 30 × 30 and 40 × 40 cm2 ), keeping the same conditions such as pressure, temperature, incremental step, direction, geometry, chamber voltage, and polarity. The PDD for 6 & 15 MV photon beams were obtained with the above mentioned field sizes and at SSD 100 cm. The obtained PDD results showed that maximum dose depth (dmax) for above all mentioned field sizes were varied within 11.8 to 15.8 mm and 21.9 to 29.9 mm respectively for 6 and 15 MV photon energies. In modern clinical radiotherapy PDD is the most essential parameters for the commissioning of medical linear accelerator. Our measured values for depth dose (dmax) and PDD applying TG-51 protocol at 10 cm depth (d10) for both 6 MV and 15 MV photon energies are found within the international refereed limits.
format Article
id 123456789-920
institution Asian University for Women
language English
publishDate 2025
publisher NUCLEAR SCIENCE AND APPLICATIONS
record_format dspace
spelling 123456789-9202026-02-18T06:15:29Z Analysis of Percentage Depth Dose for 6 and 15 MV Photon Energies of Medical Linear Accelerator With CC13 Ionization Chamber Meaze, A. K. M. M. H. Field size, percentage of depth dose, task group-51 The success of external beam radiotherapy is mainly depended on how to commission and safety precaution is maintained before beam is heated to infected cells. CC13, an intermediate size ionization chamber is frequently used for the commissioning process of dose calculation for both standard and non-standard sizes field. The measurements of PDDs (Percentage of Depth Dose) were performed using TG-51 at the North East Cancer Hospital, Bangladesh using Varian Clinac IX-5982 with 6 MV & 15 MV photon beam energies for a set of 9 field sizes (4 × 4, 6 × 6, 8 × 8, 10 × 10, 15 × 15, 20 × 20, 25 × 25, 30 × 30 and 40 × 40 cm2 ), keeping the same conditions such as pressure, temperature, incremental step, direction, geometry, chamber voltage, and polarity. The PDD for 6 & 15 MV photon beams were obtained with the above mentioned field sizes and at SSD 100 cm. The obtained PDD results showed that maximum dose depth (dmax) for above all mentioned field sizes were varied within 11.8 to 15.8 mm and 21.9 to 29.9 mm respectively for 6 and 15 MV photon energies. In modern clinical radiotherapy PDD is the most essential parameters for the commissioning of medical linear accelerator. Our measured values for depth dose (dmax) and PDD applying TG-51 protocol at 10 cm depth (d10) for both 6 MV and 15 MV photon energies are found within the international refereed limits. 2025-08-04T10:48:36Z 2025-08-04T10:48:36Z 2020 Article https://repository.auw.edu.bd/handle/123456789/920 en application/pdf application/pdf NUCLEAR SCIENCE AND APPLICATIONS
spellingShingle Field size, percentage of depth dose, task group-51
Meaze, A. K. M. M. H.
Analysis of Percentage Depth Dose for 6 and 15 MV Photon Energies of Medical Linear Accelerator With CC13 Ionization Chamber
title Analysis of Percentage Depth Dose for 6 and 15 MV Photon Energies of Medical Linear Accelerator With CC13 Ionization Chamber
title_full Analysis of Percentage Depth Dose for 6 and 15 MV Photon Energies of Medical Linear Accelerator With CC13 Ionization Chamber
title_fullStr Analysis of Percentage Depth Dose for 6 and 15 MV Photon Energies of Medical Linear Accelerator With CC13 Ionization Chamber
title_full_unstemmed Analysis of Percentage Depth Dose for 6 and 15 MV Photon Energies of Medical Linear Accelerator With CC13 Ionization Chamber
title_short Analysis of Percentage Depth Dose for 6 and 15 MV Photon Energies of Medical Linear Accelerator With CC13 Ionization Chamber
title_sort analysis of percentage depth dose for 6 and 15 mv photon energies of medical linear accelerator with cc13 ionization chamber
topic Field size, percentage of depth dose, task group-51
url https://repository.auw.edu.bd/handle/123456789/920
work_keys_str_mv AT meazeakmmh analysisofpercentagedepthdosefor6and15mvphotonenergiesofmedicallinearacceleratorwithcc13ionizationchamber